Susanne
This action may take several minutes for large corpora, please wait.
doc#118 temperature. </p><p> Another anode holder used in
doc#118 into a carbon anode holder. A
doc#118 of the carbon anode holder as indicated
doc#118 measuring the characteristic anode surface temperature (
doc#118 water-cooled anode holder were measured
doc#118 a transpiration cooled anode as well as
doc#118 a porous graphite anode cooled by a
doc#118 through a porous anode (transpiration cooling
doc#118 through the porous anode was varied systematically
doc#118 that by proper anode design the net
doc#118 column, the anode material, and
doc#118 however, the anode is still the
doc#118 air. The anode consisted of a
doc#118 anode. The anode plug (Figure
doc#118 established. The anode ablation could be
doc#118 reaction. The anode material was porous
doc#118 spacing. The anode in figure 2
doc#118 distribution along the anode holder. Three
doc#118 area at the anode by flow or
doc#118 conditions at the anode of free burning